During this measurement, the VNA firmware observes the "Low Dist" (LD) vs. "Low
Noise" (LN) gain decisions of the AGC. It then determines the statically assigned gain
for the individual sweep segments / drive ports / receivers in the following way:
if LD was selected for any of the related measurement points, then LD is assigned
otherwise LN is assigned
In other words, LN is assigned if and only if LN was selected for all related measure-
ment points.
Note:
Before running the "Learn Sweep", create the adequate port configuration and add
the required traces.
The "Learn Sweep" is not available for power sweep channels.
Remote command:
[SENSe<Ch>:]POWer:AGCMode:ACQuire
[SENSe<Ch>:]POWer:AGCMode:SAVE
Set All Items to ...
Convenience functions to apply the same manual AGC mode to all a and b wave
receivers.
If
Segmented AGC
Remote command:
[SENSe<Ch>:]POWer:GAINcontrol:ALL
5.12.4 Intermod. tab
Controls the intermodulation measurement including power calibration. The intermodu-
lation measurement requires options R&S ZNB/ZNBT-K4 and R&S ZNB/ZNBT-K14.
The results can be selected using the "Intermod." tab of the "Meas" softtool, see
ter 5.2.4, "Intermod.
The
Background information
Refer to
5.12.4.1
Controls on the Intermod tab
If
external switch matrices
Hence, except for the "Generators..." button the controls on this tab are disabled.
User Manual 1173.9163.02 ─ 62
is enabled, the setting only applies to the selected segment.
tab",
on page 300.
Intermodulation Presetting wizard
Chapter 4.7.4.3, "Intermodulation
are used, intermodulation measurements are not supported.
guides through the measurement configuration.
measurements",
on page 241.
GUI reference
Channel Config softtool
Chap-
582