R&S ZNB Series User Manual page 462

Vector network analyzers
Table of Contents

Advertisement

®
R&S
ZNB/ZNBT
Transmission normalization and "One Path Two Ports" calibration types require two-
port (Through) characterization data for the cal unit. For current calibration units, how-
ever, these two-port characterizations are typically not part of the factory characteriza-
tion. If a tooltip indicates missing two-port characterization data, simply perform a new
characterization of your cal unit. In the first dialog of the "Characterization" wizard,
select "Take All OSM and Through" to make sure that the necessary two-port data is
acquired. See also
Remote command:
[SENSe<Ch>:]CORRection:COLLect:AUTO:TYPE
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs:TYPE
[SENSe<Ch>:]CORRection:COLLect:AUTO:CONFigure
Step 2: Connections
Defines the port assignments between the R&S ZNB/ZNBT and the calibration unit.
Background information
Refer to
ple port assignments.
Figure 5-20: Multiple calibrations (and multiple port assignments)
Similar functionality as for the "SingleCal" version of the wizard (see
tions"
display and edit the corresponding port assignments.
Back
Go back to
Start
Proceeds to
If the configured port assignments are invalid, this action is disabled.
User Manual 1173.9163.02 ─ 62
"Characterization wizard"
Chapter 4.5.5.6, "Multiple port
on page 453): select the adequate calibration in the "Cal Type"/"Ports" table to
Step 1:
Ports.
Step 3: Cal
Unit.
on page 521.
assignments",
on page 191 for details on multi-
GUI reference
Cal softtool
"Step 2: Connec-
462