Calculate<Chn>:Marker<Mk>:Reference[:State]; Calculate<Chn>:Marker<Mk>:Target; Calculate<Chn>:Marker<Mk>:Reference:target; Calculate<Chn>:Marker<Mk>:Threshold - R&S ZNB Series User Manual

Vector network analyzers
Table of Contents

Advertisement

®
R&S
ZNB/ZNBT
CALCulate<Chn>:MARKer<Mk>:THReshold:STATe <Active>
CALCulate<Chn>:MARKer<Mk>:REFerence:THReshold:STATe <Boolean>
Activates or deactivates the marker-specific threshold for (single) peak searches with
the related marker (see
Use
MARKer<Mk>:REFerence:THReshold
Suffix:
<Chn>
<Mk>
Parameters:
<Active>
Manual operation:
CALCulate<Chn>:MARKer<Mk>:TYPE <Mode>
CALCulate<Chn>:MARKer<Mk>:REFerence:TYPE <Mode>
Sets the marker mode for the related marker (see
The marker must be created before using
or
ON
Suffix:
<Chn>
<Mk>
Parameters:
<Mode>
User Manual 1173.9163.02 ─ 62
"Marker addressing"
CALCulate<Chn>:MARKer<Mk>:THReshold
.
Channel number used to identify the active trace
Marker number (ignored for reference markers)
Boolean
See
CALCulate<Chn>:MARKer<Mk>:REFerence[:STATe]
.
Channel number used to identify the active trace
Marker number (ignored for reference markers)
NORMal | FIXed | ARBitrary
NORMal: If tracking is enabled
MARKer<Mk>:SEARch:TRACking ON
MARKer<Mk>:REFerence:SEARch:TRACking
er's stimulus value
CALCulate<Chn>:MARKer<Mk>:REFerence:X?) is updated
automatically with every sweep, otherwise it is fixed. The marker
position
CALCulate<Chn>:MARKer<Mk>:REFerence:Y?) is adjusted
to the corresponding response value, i.e. the marker is always
positioned on the trace.
The marker's stimulus value can be set using
CALCulate<Chn>:MARKer<Mk>:X or
CALCulate<Chn>:MARKer<Mk>:REFerence:X; the marker
automatically follows the trace.
FIXed: freezes the marker at the position determined by the cur-
rent stimulus and response value. The response value stored
with the marker is not adjusted to subsequent sweeps. Tracking
is disabled. Stimulus and response value are stored with the
marker; they are not adjusted to subsequent sweeps and trace
format changes.
on page 940) .
to set the threshold value.
"Threshold Settings"
on page 401
CALCulate<Chn>:MARKer<Mk>[:STATe]
(CALCulate<Chn>:MARKer<Mk>:X?
(CALCulate<Chn>:MARKer<Mk>:Y?
Command reference
SCPI command reference
or
CALCulate<Chn>:
"Marker addressing"
on page 940).
ON.
(CALCulate<Chn>:
or
CALCulate<Chn>:
or
ON), the mark-
or
960

Advertisement

Table of Contents
loading

Table of Contents