R&S ZNB Series User Manual page 919

Vector network analyzers
Table of Contents

Advertisement

®
R&S
ZNB/ZNBT
Parameters:
<SFDPortConfig>
Manual operation:
CALCulate:FMODel:SFD:DUT:MEASure
For a fixture modeling with the ISD tool, this command allows to measure the DUT +
Fixture at the active ports (see CALCulate:FMODel:SFD<Ph_pt>:DUT[:STATe]).
Usage:
Manual operation:
CALCulate:FMODel:SFD:DUT:MEASure:FILename <String>
For a fixture modeling with the EZD tool, this command loads the "DUT + Test Fixture"
properties from a Touchstone file.
Parameters:
<String>
Manual operation:
CALCulate:FMODel:SFD:DUT:MEASure:FILename:CLEar
Invalidates the file path previously set using
MEASure:FILename.
Usage:
Manual operation:
CALCulate:FMODel:SFD<Ph_pt>:DUT[:STATe] <Boolean>
For a fixture modeling with the SFD tool, this command allows to specify the ports to
which the test fixture is connected.
This has to be defined before measuring DUT + Fixture (using
SFD:DUT:MEASure).
Suffix:
<Ph_pt>
Parameters:
<Boolean>
Manual operation:
User Manual 1173.9163.02 ─ 62
ODD | NON
ODD: odd ports are on the left and even ports are on the right
NON: ports 1 to N are on the left and ports N+1 to 2·N are on
the right
See
"2x Thru Port Ordering"
Event
See
"Measure"
on page 632
Path to the Touchstone file, either absolute or relative to the
selected working directory (CALCulate:FMODel:DIRectory).
See
"Load File"
on page 632
CALCulate:FMODel:SFD:DUT:
Event
See
"Load File"
on page 632
.
Physical port number
1 (ON, true) if the test fixture is connected to port <Ph_pt>, 0
(OFF, false) otherwise
See
"Active"
on page 632
Command reference
SCPI command reference
on page 639
CALCulate:FMODel:
919