R&S ZNB Series User Manual page 935

Vector network analyzers
Table of Contents

Advertisement

®
R&S
ZNB/ZNBT
Example:
Manual operation:
CALCulate<Chn>:LIMit:SEGMent:COUNt?
Returns the number of limit line segments, including enabled and disabled segments.
Suffix:
<Chn>
Example:
Usage:
Manual operation:
CALCulate<Chn>:LIMit:SEGMent<Seg>:FORMula <Formula>
Defines a custom formula for limit line segment <Seg> of the related trace.
The formula is only used, if
set to ON.
Suffix:
<Chn>
<Seg>
Parameters:
<Formula>
User Manual 1173.9163.02 ─ 62
CALC:LIM:DATA 1,1500000000, 2000000000,2,3
Define an upper limit line segment (segment no. 1) in the stimu-
lus range between 1.5 GHz and 2 GHz, assigning response val-
ues of +2 dB and +3 dB.
:CALC:LIM:SEGM:AMPL:STAR 5; STOP 5; :CALC:LIM:
SEGM:TYPE LMIN
Change the segment to a lower limit line segment with a con-
stant response value of +5 dB.
CALC:LIM:DATA?
Query the type, the stimulus and response values of the created
segment with a single command. The response is
2,1500000000,2000000000,5,5.
See
"Segment List"
on page 378
.
Channel number used to identify the active trace
CALC:LIM:DATA 1,1500000000, 2000000000,2,3
Define an upper limit line segment (segment no. 1) in the stimu-
lus range between 1.5 GHz and 2 GHz, assigning response val-
ues of +2 dB and +3 dB.
CALC:LIM:SEGM:COUNT?
Query the number of segments. The response is 1.
Query only
See
"Segment List"
on page 378
CALCulate<Chn>:LIMit:SEGMent<Seg>:FORMula
.
Channel number used to identify the active trace
Segment number
User-defined formula (string).
Allows for the same mathematical expressions as
CALCulate<Chn>:MATH[:EXPRession]:SDEFine.
Command reference
SCPI command reference
935
is

Advertisement

Table of Contents
loading