R&S ZNB Series User Manual page 478

Vector network analyzers
Table of Contents

Advertisement

®
R&S
ZNB/ZNBT
Remote command:
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]:VERification:
RESult?
Port Overview
Shows all source ports together with the possible power calibrations.
For each analyzer port P1 ... PN, a "Ref. Receiver", a "Meas. Receiver" and a "Source
Flatness" calibration can be performed.
If a "SMARTerCal" is active, additional receiver power calibrations are not allowed
and the R&S ZNB/ZNBT only offers the possibility to perform "Source Flatness"
calibrations of the related test ports:
See
on page 200 for details.
If external generators are configured, they appear as additional source ports G1 ...
Gm in the port overview. A source power (flatness) calibration is available at each
generator port.
Select a "Ref. Receiver", "Meas. Receiver", or "Source Flatness" symbol to start the
related calibration. See
Cal dialog – Meas. Receiver"
on page 481.
A green checkmark
fully. A half circle overlay
ered (see
Remote command:
SOURce<Ch>:POWer:CORRection[:ACQuire]
[SENSe<Ch>:]CORRection:POWer<PhyPt>:ACQuire
SOURce<Ch>:POWer:CORRection:DATA:PARameter<Wv>?
SOURce<Ch>:POWer<PhyPt>:CORRection:DATA:PARameter<Wv>:COUNt?
Apply
Is enabled when a new set of power calibration data has been acquired. The button
applies all available power calibrations to the active channel, aborts the verification
sweeps, and closes the port overview section.
The power calibration state is indicated in the trace list, see
calibration
on page 531 to activate, deactivate, or store power calibrations.
Remote command:
SOURce<Ch>:POWer:CORRection[:ACQuire]
[SENSe<Ch>:]CORRection:POWer<PhyPt>:ACQuire
SOURce<Ch>:POWer:CORRection:DATA
User Manual 1173.9163.02 ─ 62
Chapter 4.5.7.3, "Combining SMARTerCal with scalar power
"Power Cal dialog – Ref. Receiver"
on page 480, and
indicates that the calibration data has been acquired success-
indicates that only the respective port frequencies are cov-
"Calibrate Only Port Frequency"
labels",
on page 197. Use the functions in the
on page 479, and
"Power Cal dialog – Source Flatness"
on page 480).
Chapter 4.5.6.3, "Power
Chapter 5.11.4, "Use Cal
GUI reference
Cal softtool
calibration",
"Power
tab",
478

Advertisement

Table of Contents
loading