Remote command:
CALCulate<Chn>:STATistics:NLINear:COMP[:STATe]
CALCulate<Chn>:STATistics:NLINear:COMP:LEVel
CALCulate<Chn>:STATistics:NLINear:COMP:PHASe
CALCulate<Chn>:STATistics:NLINear:COMP:RESult?
CALCulate<Chn>:STATistics[:STATe]:AREA
Reference Value
This control is only available if
reference value ("small signal value") for the compression point calculation.
"First Point"
(default)
"Selected
Marker"
"Range"
"Defined
Value"
Remote command:
CALCulate<Chn>:STATistics:NLINear:COMP:REFerence
Selected Marker
This button is only visible if "Selected Marker" is used as
you to select the marker whose value shall be used as the reference ("small signal
value") for the compression point calculation.
Remote command:
CALCulate<Chn>:STATistics:NLINear:COMP:RMARker
Ref. Range
This button is only visible if "Range" is used as
It opens the "Reference Range" dialog that allows you to configure the sweep range
whose average value shall be used as the reference ("small signal value") for the com-
pression point calculation.
User Manual 1173.9163.02 ─ 62
compression point
uses the value at the first point of the selected evaluation range as
the reference value (see
on page 360)
uses the value at a selectable marker as the reference value (see
"Selected Marker"
on page 359)
uses the average value in a configurable
as the reference value
allows to specify the reference value manually (see
on page 360)
statistics are calculated. It selects the
Chapter 5.5.8.2, "Evaluation Range
"Ref. Range"
Reference
Reference
Value.
GUI reference
Trace Config softtool
dialog",
on page 359
"Defined Value"
Value. It allows
359