R&S ZNB Series User Manual page 980

Vector network analyzers
Table of Contents

Advertisement

®
R&S
ZNB/ZNBT
Suffix:
<Chn>
Parameters:
<Boolean>
Example:
Manual operation:
CALCulate<Chn>:RIPPle:FAIL?
Returns a 0 or 1 to indicate whether or not the global ripple limit check has failed.
Tip: Use
for a single ripple limit range.
Suffix:
<Chn>
Example:
Usage:
Manual operation:
CALCulate:RIPPle:FAIL:ALL? [<RecallSet>]
Returns a 0 or 1 to indicate whether or not the global ripple limit check has failed for at
least one channel in the referenced recall set.
Query parameters:
<RecallSet>
Usage:
Manual operation:
User Manual 1173.9163.02 ─ 62
.
Channel number used to identify the active trace
ON | OFF - ripple limit line on or off.
*RST:
*RST; CALC:RIPP:DATA 1,1500000000, 2000000000,
3
Define and enable a ripple limit range in the stimulus range
between 1.5 GHz and 2 GHz, assigning a ripple limit of +3 dB.
CALC:RIPP:DISP ON
Show the ripple limit range in the active diagram.
See
"Show Ripple Limits"
CALCulate<Chn>:RIPPle:SEGMent<Seg>:RESult?
.
Channel number used to identify the active trace
*RST; CALC:RIPP:DATA 1, 1500000000, 2000000000,
3
Define and enable a ripple limit range in the stimulus range
between 1.5 GHz and 2 GHz, assigning a ripple limit of +3 dB.
CALC:RIPP:STAT ON; FAIL?
Switch the limit check on and query the result.
CALC:RIPP:STAT:AREA LEFT, TOP
For a subsequent check at the GUI or a hardcopy, move the info
field to the top left position.
Query only
See
"Ripple Check"
Recall set name; if omitted the active recall set is used
Query only
See
"Ripple Check"
OFF
on page 382
on page 383
on page 383
Command reference
SCPI command reference
to query the result
980

Advertisement

Table of Contents
loading