R&S ZNB Series User Manual page 383

Vector network analyzers
Table of Contents

Advertisement

®
R&S
ZNB/ZNBT
Ripple Check
Switches the ripple limit check of the active trace on or off. When the limit check is
switched on, a movable info field shows the pass/fail information and the measured rip-
ple in each ripple limit range.
Limit violations are marked with squares, whose color is defined in the
Color Scheme dialog
A
TTL signal
Note:
Ripple check and display of limit lines are independent of each other:
For each trace, ripple limits can only be set for a single cartesian trace format. If
another trace format is selected, the ripple limit lines are hidden and the ripple
check is suspended.
Remote command:
CALCulate<Chn>:RIPPle:STATe
CALCulate<Chn>:RIPPle:FAIL?
CALCulate:RIPPle:FAIL:ALL?
CALCulate<Chn>:RIPPle:SEGMent<Seg>:RESult?
CALCulate<Chn>:RIPPle:STATe:AREA
Show Results All Traces
Defines the visibility of ripple info fields in the active recall set.
If disabled, only the ripple info field of the active trace is displayed (in case it has
ripple check enabled).
If enabled, ripple info fields are displayed for all traces with ripple check enabled.
Remote command:
CALCulate:RIPPle:DISPlay:RESult:ALL[:STATe]
Clear Test
Resets the limit check results.
User Manual 1173.9163.02 ─ 62
("Limit Fail Trace Color").
indicating pass or fail can be generated in addition.
The ripple limits can be displayed, no matter if the ripple check is enabled.
If "Ripple Check" is enabled, the ripple limits are checked, no matter if they are
displayed.
The ripple check can even be enabled, if no limit lines are defined. In this case,
the info field displays "No ripple defined!" and the limit check always passes.
GUI reference
Lines softtool
Define User
383

Advertisement

Table of Contents
loading

Table of Contents