R&S ZNB Series User Manual page 199

Vector network analyzers
Table of Contents

Advertisement

®
R&S
ZNB/ZNBT
B: Two-port at power meter (during calibration)
Test and measurement procedure:
1. Perform the calibration with the additional two-port between the analyzer port and
the power sensor. During the calibration, the analyzer increases the power sensor
values by the 2-port transmission coefficients to move the calibration plane of the
power calibration towards the input of the DUT. The calibration plane corresponds
to the input of the additional 2-port.
2. Perform the measurement without the additional two-port.
Practical example: An adapter or attenuator with known attenuation is needed to con-
nect the power sensor to the test port of the network analyzer. The transmission coeffi-
cients of the adapter are used for the power meter correction.
4.5.7 SMARTerCal
A SMARTerCal (smarter calibration) is a combination of a "One Path Two Ports" or a
full n-port system error correction (TOSM, UOSM, Adapter Removal, TRL, TNA ...) for
two or more ports with a (scalar) receiver power calibration at a single port. The two
calibration methods serve different purposes:
The system error correction requires a set of calibration standards; it provides vec-
tor error-corrected S-parameters. For equal port frequencies, the n-port calibration
types provide the full set of error terms. For frequency conversion measurements,
a source match correction and (optional) load match correction is calculated.
The receiver power calibration requires an external power meter; it corrects the
power readings of the reference and measurement receivers according to the mea-
sured absolute power at the calibration plane. This does not include a readjust-
ment of the actual source power (flatness calibration).
Example: Channel base power: –10 dBm; the test setup involves a 3-dB attenua-
tion between the source port and the calibration plane. After the power calibration
is applied, the analyzer indicates an output power (a-wave) of –13 dBm, although
the actual source power remains at –10 dBm.
The SMARTerCal is also applied to ratios and wave quantities. For measurements on
linear DUTs, SMARTerCal is sufficient. Non-linear measurements can be further
improved by a combination of a SMARTerCal plus a scalar power calibration. See
Combining SMARTerCal with scalar power
User Manual 1173.9163.02 ─ 62
Concepts and features
calibration.
Calibration
199

Advertisement

Table of Contents
loading

Table of Contents