R&S ZNB Series User Manual page 439

Vector network analyzers
Table of Contents

Advertisement

®
R&S
ZNB/ZNBT
Remote command:
TRIGger<Ch>[:SEQuence]:SOURce MANual
*TRG
Sequence
Selects the measurement cycle or sequence of actions to be triggered in "External" or
"Manual" mode.
"Sweep" means that each trigger event starts an entire sweep, according to the
current sweep configuration.
"Point" means that each trigger event starts the measurement at the next sweep
point.
"Partial Measurement" means that each trigger event starts the next partial mea-
surement at the current or at the next sweep point. If every sweep point only
requires a single partial measurement, this option is equivalent to "Point".
See also
"Segment" means that each trigger event starts the next sweep segment within the
current sweep. If a sweep type other than
alent to "Sweep".
Relation with other sweep settings
Some sweep settings are logically incompatible with a particular selection of the trig-
gered measurement sequence:
If a
"Alternated"
sequence comprises more than one sweep point. If "Point" or "Partial Measure-
ment" is selected, "Alternated" mode is switched off and vice versa.
Note: The trigger events must be adjusted to the triggered measurement sequence. If
the analyzer receives a trigger event while the last sequence is still running, the
R&S ZNB/ZNBT skips the trigger event and generates a message.
Remote command:
TRIGger<Ch>[:SEQuence]:LINK
Delay
Specifies a delay time between the trigger event and the start of the next measurement
sequence.
The specified "Delay" must be zero or positive, so that the trigger event precedes the
start of the measurement (post-trigger).
If "Multiple Triggers" is active, the "Delay" can be selected individually using the
Manager
User Manual 1173.9163.02 ─ 62
Chapter 4.1.4.1, "Partial measurements and driving
Time
sweep is performed, the sequence is always a sweep.
Driving Mode
dialog.
Segmented
only makes sense if the triggered measurement
GUI reference
Sweep Softtool
mode",
on page 91.
is active, this option is equiv-
Trigger
439