Remote command:
CALCulate<Chn>:RIPPle:CLEar
Global Check
See
TTL1 Pass / TTL2 Pass
See
5.6.2.2
Define Ripple Test dialog
The "Define Ripple Test" dialog defines the ripple limits for the active trace on a range-
by-range basis. A separate ripple limit can be assigned to each range.
Access: Trace – [Line] > "Ripple Test" > "Def. Ripple Test..."
Defining ripple limits with minimum effort
Choose one of the following methods to create and handle ripple limit ranges effi-
ciently:
To configure a limit test with only a few ranges, use "Add" and edit each range in
the table individually.
Use the "Align All" button to create non-overlapping, contiguous ranges of equal
width.
Save your ripple ranges to a file so you can reuse or modify them in later sessions
("Save Ripple Test..., Recall Ripple Test...").
Background information
Refer to
The "Define Ripple Test" dialog contains a table to edit the individual ranges of the rip-
ple check ranges. The buttons below the table extend, shorten, or reorder the range list
and save/recall ripple test data.
Range List
Defines the individual ripple limit ranges.
User Manual 1173.9163.02 ─ 62
"Global Check"
on page 376.
"TTL1 Pass / TTL2 Pass"
Chapter 4.4.1, "Limit
on page 376.
check",
on page 144.