Direct Compensation
With "Direct Compensation", a frequency-dependent transmission factor is calculated;
see
"Auto Length and Loss vs. Direct Compensation"
Remote command:
[SENSe:]CORRection:COLLect:FIXTure:LMParameter[:STATe]
[SENSe<Ch>:]CORRection:OFFSet<PhyPt>:DFComp[:STATe]?
Prompt for Each Port
Determines how the R&S ZNB/ZNBT performs the sweeps for a given termination type
(Open and/or Short; see
If unchecked, it performs the sweeps for Open/Short without interruption, implicitly
assuming that all ports are terminated accordingly
If checked, it interrupts the data acquisition process after each port, which allows
you to modify the test setup (e.g. terminate the next measured port).
Measurement Type
The "Open" , "Short", and "Open and Short" buttons bring up the
ard" dialog that guides you through the actual fixture measurement. See
vs. Open and Short compensation"
Measure Fixture wizard
The "Measure Fixture" dialog guides you through the previously configured fixture
compensation measurements.
Figure 5-56: Measure Fixture dialog: Auto Length (and Loss)
To acquire the necessary data, proceed as indicated in the information area.
With Prompt for Each Port disabled, "Take" acquires data for all selected ports in one
go. Otherwise sweeps are taken port by port.
For "Direct Compensation", it is also possible to save the acquired data to file. In future
measurements, you can load these files instead of repeating the data acquisition.
User Manual 1173.9163.02 ─ 62
"Measurement Type"
on page 620).
on page 207.
GUI reference
Offset Embed softtool
on page 207.
"Measure Fixture wiz-
"Open/Short
620