R&S ZNB Series User Manual page 772

Vector network analyzers
Table of Contents

Advertisement

®
R&S
ZNB/ZNBT
Selftest page
In case you selected "Selftest" on the
Check" wizard is the standard
Execute/Repeat
Executes the selftest for the first time in this "Operators Check" wizard session, or
repeats it.
Remote command:
SYSTem:OPCHeck:SELFtest?
Power Check page
In case you selected "Power Check" on the
port power checks from this page.
During the port power check, the firmware performs a 201-point-linear frequency
sweep over the whole frequency range of the analyzer.
By default, the source power is set to -10 dBm and the firmware checks whether the
measured a- and b-waves are at most 3 dBm above or below this value.
User Manual 1173.9163.02 ─ 62