R&S ZNB Series User Manual page 850

Vector network analyzers
Table of Contents

Advertisement

®
R&S
ZNB/ZNBT
CALCulate<Ch>:DATA:CALL <Format>, <Data>...
The query reads the current response values of all S-parameter data traces at channel
<Ch>'s data access point <DACPoint>.
If a full n-port system error correction is active in the related channel, the command
reads the full nxn S-matrix of the calibrated ports (there is no need to create or display
the S-parameter traces). Use
available traces.
The "set direction" allows you to import measurement data at the "Fixture Simulation
Input" data access point (see
Note
Importing data is only supported in single sweep mode
CONTinuous
Before importing data, the channel must be configured with the same settings that
were used during export (user calibration, balanced port configuration, stimulus
axis etc.).
Furthermore the switch matrix configuration must match.
After importing data, "downstream" parameters in the data flow can be changed
and their effect is shown directly
Any request for new data from the hardware (Restart Sweep in single sweep mode
or switching to continuous sweep mode) discards imported data and switches back
the display to measured data
Suffix:
<Ch>
Parameters:
<Format>
<Data>
Example:
User Manual 1173.9163.02 ─ 62
CALCulate<Ch>:DATA:CALL:CATalog?
Chapter 4.1.5, "Data
OFF)
.
Channel number
SDATa | FSIData
SDATa
Output as unformatted trace data; see
. Query only.
FSIData
Output or input at "Fixture Simulation Input" data access point.
Trace data either in ASCII or block data format, depending on
the current
FORMat[:DATA]
The column order must match the one used during export.
Suppose that a TOSM calibration for ports 1 and 2 is active in
channel no. 1.
CALCulate:DATA:CALL:CATalog?
Query the traces available for CALCulate<Ch>:DATA:CALL?.
The response is 'S11,S12,S21,S22'.
CALCulate:DATA:CALL? SDATa
Return the complex response values of all traces. The traces in
the catalog list are read one after another: The response array
contains n (number of points) pairs of real and imaginary values
for S
, followed by n pairs of values for S
11
Command reference
SCPI command reference
flow",
on page 98).
(INITiate<Ch>:
CALCulate<Chn>:DATA
setting.
, S
, and S
12
21
to query the
.
22
850

Advertisement

Table of Contents