R&S ZNB Series User Manual page 350

Vector network analyzers
Table of Contents

Advertisement

®
R&S
ZNB/ZNBT
Note: The frequency domain window is used to filter the trace before transforming it to
the time domain. An independent "Time Gate" can be used after the transformation to
eliminate unwanted responses (see
The analyzer always uses a "No Profiling (Rectangle)" window to calculate the time-
gated frequency domain trace, see
Remote command:
CALCulate<Chn>:TRANsform:TIME:WINDow
Side Lobe Level
Defines the side lobe suppression for an "Arbitrary Sidelobes (Dolph-Chebychev)" win-
dow. The entered value is the ratio of the power of the central lobe to the power of the
first side lobe in dB.
Remote command:
CALCulate<Chn>:TRANsform:TIME:DCHebyshev
Resolution Enh.
Broadens the frequency range that the analyzer considers for the time domain trans-
form by a linear factor. A factor of 1 means that the original sweep range and the mea-
sured sweep points are used; no additional assumptions are made. With higher resolu-
tion enhancement factors, the measurement data is extrapolated using a linear predic-
tion method. As a result, the resolution in time domain can be improved.
The ideal resolution enhancement factor depends on the properties of the DUT. For
distance to fault measurements on cables, set it to 1.
Remote command:
CALCulate<Chn>:TRANsform:TIME:RESolution:EFACtor
TD-VSWR
Enables time domain site VSWR measurements (see
S
VSWR
Remote command:
CALCulate:TDVSwr[:STATe]
Gate Span
Time domain S
na's direct response (plus ring-down time), separating the direct response from the
indirect responses (reflections).
"Gate Span" is equivalent to the "Span" value on the
This button is only enabled if
Remote command:
CALCulate<Chn>:FILTer[:GATE]:TIME:SPAN
Show Impulse Response...
To get an impression of the required
ing the impulse response of the active trace in a new diagram.
This button is only enabled if
User Manual 1173.9163.02 ─ 62
measurements",
on page 228).
measurements
VSWR
Chapter 5.5.6, "Time Gate
"Time-gated frequency domain trace"
relies on a time gate that is centered at the anten-
TD-VSWR
is active.
Gate
Span, you can quickly create a trace display-
TD-VSWR
is active.
GUI reference
Trace Config softtool
tab",
on page 353).
on page 228.
Chapter 4.7.2.6, "Time domain
Time Gate
tab.
350

Advertisement

Table of Contents
loading

Table of Contents