R&S ZNB Series User Manual page 754

Vector network analyzers
Table of Contents

Advertisement

®
R&S
ZNB/ZNBT
Auto Power Setting for Cal Unit
Sets the source power at all test ports to -10 dBm while an automatic calibration is
active. Applying this source power to the ports of the calibration unit ensures best
accuracy of the automatic calibration. The source powers are reset to their original val-
ues after the calibration is completed. The automatic power reduction can be deactiva-
ted in case that the test setup introduces a large attenuation.
Remote command:
SYSTem:COMMunicate:RDEVice:AKAL:PREDuction[:STATe]
Auto Averaging
Activates automatic averaging, which means that the VNA performs multiple calibration
sweeps and applies averaging to reduce trace noise. In contrast to regular averaging
(see
calculated automatically.
Remote command:
[SENSe:]CORRection:COLLect:AVERage
Multiple Cal in Calibration Wizard
Enables/disables multiple calibrations in the calibration wizard, which allows to set up
and perform multiple calibrations in one manual calibration procedure. See
ter 5.11.1.4, "Calibration Unit wizard
bration Setting wizard
(Cal Unit,
ual,
MultiCal)",
Remote command:
SYSTem:DISPlay:DIALogs:SETup:MCAL[:STATe]
Show Cal Kit Label
Enables/disables the "Calibration Info" dialog during manual calibration (see
Sweep"
Independent of the state of the "Show Cal Kit Label" flag, cal kit labels are displayed in
several other manual calibration dialogs.
Use Reduced Number of Through
Enables or disables the
matic calibration.
Note that for this to work, the port assignments must have a star topology, with a
"Common Port" set using
on page 1219.
Remote command:
[SENSe:]CORRection:COLLect:AUTO:MCONnect
Delete Cal Pool / Delete All Cal Kits
Deletes all calibration data and all cal kit data. See
ager
Remote command:
n/a
User Manual 1173.9163.02 ─ 62
Chapter 5.9.3, "Average
(MultiCal)",
MultiCal)",
on page 491 and
on page 505).
on page 470).
Reduced Through
[SENSe<Ch>:]CORRection:COLLect:AUTO:CPORt
dialog",
on page 534.
tab",
on page 421), the number of calibration sweeps is
(MultiCal)",
on page 458,
on page 472,
Chapter 5.11.1.9, "SMARTerCal wizard
Chapter 5.11.1.11, "SMARTerCal wizard (man-
logic for every port assignment of an auto-
Chapter 5.11.4.3, "Calibration Man-
GUI reference
Setup softtool
Chap-
Chapter 5.11.1.6, "Cali-
"Start Cal
754