ZNB/ZNBT
4 paths per transmission measurement for an N-port VNA with N≥4
This kind of "multipath calibration" offers the following additional benefit:
During manual calibration, the measured reflection/transmission coefficients are pre-
sented as memory traces – one per path. By comparing these traces, it should be pos-
sible to track down hardware problems (cables, connectors, matrix, ...) already during
calibration.
4.7.30 External DLLs
The R&S ZNB/ZNBT firmware provides a plug-in interface that allows you to extend
the VNA firmware with custom functionality (measurements, device control, ...) .