If the "Fixture Simulator" is disabled for the related channel (see
on page 549), this tab is inactive, i.e. all controls except the "Overview" button are
grayed out.
Overview
See
Type
Switches between "Deembedding" and "Embedding" network definition.
All Deembedding Activated
Deactivates or activates all deembeddings (single-ended, balanced, port set, ground
loop) for the active channel, if any.
This control is only visible if
Network
The graphical list contains all available 1-port networks:
The
The "1-Port Data" network is defined by imported S-parameter data; see
The remaining networks (Shunt L and Shunt C) are defined by lumped elements
whose parameters are displayed below the graphical list.
Tip:
Drag and drop the network symbols in horizontal or vertical direction to switch to
the next symbol.
Use the
in "Capacitor in parallel with resistor" model to conductances and vice versa.
User Manual 1173.9163.02 ─ 62
Chapter 4.6.2.7, "Ground loop
"Overview"
on page 616.
symbol selects "no network" and disables ground loop de-/embedding.
Conductance in Embedding Networks
de-/embedding",
Type
= "Deembedding".
GUI reference
Offset Embed softtool
on page 216.
"Fixture Simulator"
Select File...
switch to change from resistances
647