ZNB/ZNBT
Suffix:
<Ch>
Setting parameters:
<TraceName>
<Result>
Manual operation:
CALCulate<Ch>:PARameter:SDEFine <TraceName>, <Result>
Creates a trace and assigns a channel number, a name and a measurement parame-
ter to it. The trace becomes the active trace in the channel but is not displayed.
Note: To display the trace defined via CALCulate<Ch>:PARameter:SDEFine, cre-
ate a diagram
diagram (DISPlay[:WINDow<Wnd>]:TRACe<WndTr>:FEED); see example below.
CALCulate<Ch>:PARameter:MEASure
existing trace.
To select an existing trace as the active trace, use
SELect. You can open the trace manager to obtain an overview of all channels and
traces, including the traces that are not displayed.
Tip: This command has no query form. Use
<TraceName> to query the measurement result of the trace.
PARameter:CATalog?
Suffix:
<Ch>
Setting parameters:
<TraceName>
<Result>
Example:
User Manual 1173.9163.02 ─ 62
.
Channel number
Trace name
Measured quantity
See
"S-Parameter (selector)"
(DISPlay[:WINDow<Wnd>][:STATe]
returns a list of all defined traces.
.
Channel number referencing a previously defined channel. If
<Ch> does not exist, it is generated with default channel set-
tings.
Trace name, string variable, e.g. 'Trc4'. See "Rules for trace
names" in
"Table Area"
unique across all channels and diagrams.
If a trace with the selected trace name already exists, the ana-
lyzer behaves as follows: If the existing trace is assigned to the
same channel as the new trace, it is deleted. The new trace is
not automatically assigned to a diagram area; see note above.
If the existing trace is assigned to a different channel, no new
trace can be created. The analyzer returns an error message.
Measurement result (string variable), see table below.
CALC4:PAR:SDEF 'Ch4Tr1', 'S11'
Create channel 4 and a trace named Ch4Tr1 to measure the
input reflection coefficient S
DISP:WIND2:STAT ON
Create diagram area no. 2.
DISP:WIND2:TRAC:FEED 'CH4TR1'
Display the generated trace in diagram area no. 2.
Command reference
SCPI command reference
on page 285
ON) and assign the trace to this
changes the measurement result of an
CALCulate<Ch>:PARameter:
CALCulate<Ch>:PARameter:MEASure
CALCulate<Ch>:
on page 337. Trace names must be
.