ZNB/ZNBT
Chapter 5.13.5.8, "EZD Advanced
Use Impedance Correction
Enables/disables the impedance correction functionality of the respective tool.
This is a global setting.
Remote command:
CALCulate:FMODel:EZD:IMPedance
CALCulate:FMODel:ISD:IMPedance
CALCulate:FMODel:SFD:IMPedance
(CALCulate:FMODel:SFD:AUTO)
Suppress Remeasure Warnings
If checked, the
ings, if the measurement is repeated with different ports.
This is a global setting.
Remote command:
n.a.
Reset to Default
Restores the default settings of the selected fixture modeling tool. This comprises:
The states and settings in the
Save Fixture Model
The global settings of the selected tool (tool specific)
Remote command:
CALCulate:FMODel:ISD:PRESet
CALCulate:FMODel:SFD:PRESet
CALCulate:FMODel:EZD:PRESet
Apply
This button is only active, if the selected deembedding tool has been successfully run
(see
Applies the calculated deembedding files <...>_left_DUT.sNp and
<...>_right_DUT.sNp) to the ports marked "Active" in the
tion.
Remote command:
MMEMory:LOAD:VNETworks<Ch>:SENDed:DEEMbedding<PhyPt>
MMEMory:LOAD:VNETworks<Ch>:BALanced:DEEMbedding<LogPt>
Cancel
Closes the dialog without applying any deembedding files.
5.13.5.2
Measure Coupon
The "Measure Coupon" section allows you to measure one or more test coupons for
the related fixture.
User Manual 1173.9163.02 ─ 62
Measure Coupon
sections (common for all tools)
Chapter 5.13.5.4, "Save Fixture
Settings",
on page 639
and
Measure DUT + Test Fixture
Measure
Coupon,
Measure DUT + Test Fixture
Model",
on page 632).
GUI reference
Offset Embed softtool
do not raise warn-
and
Save Fixture Model
sec-
629

Advertisement

Table of Contents
loading

Table of Contents