GE Healthcare
Revision 1
Job Card CFG A001 - Software Reload
•
Follow the Install from scratch instructions given in the booklet supplied with the AWS Application
Software CD-ROM to install the software. Before starting the procedure, take note of the require-
ments for Material needed.
•
After completing the installation, Refer to
771:
-
Restore AWS parameters from the system backup on the backup media (either floppy diskette or
CD-R). If this backup media is not available, the procedures given in the
AWS Configuration on page 573
-
Back up the system configuration to a floppy diskette or a CD-R.
! Notice:
If the system backup is not done, no backup for IDC and Senographe parameters will exist.
•
For V3 Control Stations that use AWS V53.10.10 or later and V4 Control Stations, install the Embed-
ded Publications, see
•
For Senographe systems that use AWS V43.0 or later (on the SB150) or AWS 53.0 or later (on the
U20) or AWS 54.10 or later (on the Z400), if the Senographe system does not use the default AOP
Classic (AOP A) tables configuration, restore as necessary the AOP configuration (see
ELE A031 - AOP Configuration on page
6-2
AWS Application Reload or Update
•
To reload the Application software only, or if a new version of application software must be installed
on your system, use the AWS Application Software CD-ROM used for the current installation, or that
provided for the update, and follow the Update from CD-ROM instructions given in the CD-ROM
booklet.
•
For V3 Control Stations that use AWS V53.10.10 or later and V4 Control Stations, reinstall the
Embedded Publications, see
•
For Senographe systems that use AWS V43.0 or later (on the SB150) or AWS 53.0 or later (on the
U20) or AWS 54.10 or later (on the Z400), if the Senographe system does not use the default AOP
Classic (AOP A) tables configuration, restore as necessary the AOP configuration (see
ELE A031 - AOP Configuration on page
7
IDC SOFTWARE
7-1
IDC Software Reload or Update from CD-ROM
Software reload from CD-ROM is necessary when the IDC Computer has been replaced, or if a new ver-
sion of the IDC software must be installed on your system. Use the following steps to perform an IDC
Software Reload:
1. Back up the system configuration (including the IDC configuration) to the AWS hard disk. Refer to
Job Card ELE A052 - Back-up/Restore Procedure on page
2. Follow the installation instructions given on the back of the CD-ROM case of the IDC Application
Software CD-ROM used for the current installation or supplied for a software update.
3. Restore the IDC directory data from the current system backup from the AWS hard disk. Refer to
Card ELE A052 - Back-up/Restore Procedure on page
4. Reload the detector paramaters according to
D/R A141 - Image Receptor/Digital Detector
eters for the detector).
Chapter 9
Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A052 - Back-up/Restore Procedure on page
must be carried out.
Job Card ELE A044 - Install User Publications on page
693).
Job Card ELE A044 - Install User Publications on page
693).
Page no. 1642
771.
771.
Load Detector Parameters on page 1103
(requires the floppy disk or CD-ROM containing param-
Senographe DS
Job Card ELE A013 -
755.
Job Card
755.
Job Card
Job
in
Job Card
JC-CFG-A-001.fm
Need help?
Do you have a question about the Senographe DS and is the answer not in the manual?