Job Card D/R A287 - Collimator DC-DC Board
TP17
5. Power on the Gantry and check that no errors are raised on the Gantry LCD display.
6. Check that the LED states are as follows. For more information about the behavior and states of the
Collimator Control board LEDs, refer to
7. Reboot the Senographe system and verify that the Collimator blades move as expected.
7
COMPLETION
None.
Chapter 9
Service Information and Procedures Class A 2385072-16-8EN
Collimator Control Board PL403 on page
LED Label
DS1
DS2
DS3
DS4
DS5
DS6
DS7
DS8
DS9
DS10
LED Color
Green
Green
Green
Green
Green
Green
Red
Red
Red
Red
Page no. 1382
Senographe DS
295.
LED Status
ON
OFF or ON
ON
ON
ON
OFF
OFF
OFF
OFF
OFF
TP20
CC1TP
JC-DR-A-287.fm