Revision 1
Job Card D/R A232 - Lift Screw Assembly
ILLUSTRATION 1 - EXAMPLES OF GOOD AND BAD GREASING OF THE LEADSCREW
2. Once you have finished applying grease to the entire Leadscrew, use the slow speed Arm Control
buttons to drive the Lift over the whole range up and down two times in slow speed.
3. Use the fast speed Arm Control buttons to drive the Lift over the whole range up and down a few
times in fast speed.
4. Check how grease has been spread to the screw. Examples of well spread grease, smoothly and
uniformly along the whole screw length, with no splattered grease are shown in illustration 2.
Chapter 9
Service Information and Procedures Class A 2385072-16-8EN
Bad
Too much applied grease can splatter to column
wall. Clean surfaces, walls and cables from spat-
tered grease.
Page no. 1276
Senographe DS
Good
Uniformly applied grease.