Job Card D/R A247 - Motorized Compression Brake - GE Senographe DS Manual

Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1

Job Card D/R A247 - Motorized Compression Brake

1
SUPPLIES
Motorized Compression Brake
2
TOOLS
Standard Tool Box
Mammography compression scale with digital display
3
REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 45 minutes
4
SAFETY PRECAUTIONS
WARNING
The Gantry must be powered on, mark it clearly with a warning sign to prevent others from
applying commands.
5
PREREQUISITES
Rotate the arm to 90° for access to the Motorized
Compression Brake (1).
Switch off the Gantry power.
Remove the Compression Arm covers (refer to
Card PHY A044 - Remove/Reinstall Gantry Covers
on page
523).