Job Card D/R A247 - Motorized Compression Brake
6
PROCEDURE
6-1
Disassembly of the Motorized Compression Brake
1. Loosen the screw (2) on the top of the brake (2.5 mm
allen wrench)
2. Disconnect the brake power supply cable (3) (W309) from the Compression Board (connector J5)
3. Connect the brake power supply cable (3) (W309) to the Arm Distribution Board (connector J5) in
place of the previous cable (4) (disconnect W307).
Compression Board
Chapter 9
Service Information and Procedures Class A 2385072-16-8EN
3
Connect W309
Here
4
Page no. 1310
Senographe DS
2
Arm Distribution Board