GE Healthcare
Revision 1
Job Card D/R A321 - IDC Computer
6. For V2 Control Stations:
a. disconnect the front two allen (3) screws that attach the IDC Computer to the Control Station
framework (2.5 mm allen wrench).
b. Disconnect the back two allen (4) screws that attach the IDC Computer to the Control Station
framework (2.5 mm allen wrench).
7. Remove the defective IDC Computer from the Control Station.
Chapter 9
Service Information and Procedures Class A 2385072-16-8EN
3
Page no. 1506
Senographe DS
4
JC-DR-A-321_DS.fm