Reassembly of New Mouse on Sisfle Arm
1. Place the new mouse at the back of the mouse support (so that there is enough cable for use of the
mouse on the entire mouse support).
2. Use pointed nose pliers to create an opening in the harness at 27 cm from the end of the harness.
3. Route the mouse cable through the opening in the harness, so that it leaves the other end of the har-
ness. Then route the mouse cable through the rotative arm support entrance into the Control Station.
4. Ensure that the rotative arm (1) and rotative arm support (2) are aligned in the same plane. Carefully
put the monitor harness inside the rotative arm and rotative arm support.
JC-DR-A-312.fm
Service Information and Procedures Class A 2385072-16-8EN
1
Page no. 1423
Senographe DS
Job Card D/R A312 - Mouse
27 cm
Aligned in the same plane.
2