7
COMPLETION
7-1
Firmware Check
Check whether the new Tube Tilt Board firmware needs to be updated, as follows:
1. Reboot the Senographe system.
2. Go to the Service Desktop home page (see
3. Consult Table 1 on page 63 in the section
on page
63. Determine the system level from the ADS/IDC/Generator CPU versions displayed on the
Service Desktop home page, and by consulting Table 1 on page 63. The system level will be defined
by one of the System Level (Model Number) columns in Table 1 on page 63.
4. Check whether the current firmware version(s) for the new Tube Tilt Board corresponds to the
required firmware version for the that system level.
If the current firmware version(s) match the required version(s) quoted in Table 1 on page 63, no
additional steps are required.
If the current firmware version(s) does not match the required version(s), you must perform a firm-
ware update of the mismatched firmware version(s) so that the new Tube Tilt Board firmware ver-
sion(s) match those quoted in Table 1 on page 63. To update the new Tube Tilt Board firmware
follow the instructions supplied in the Gantry Software Update CD-ROM that is kept with the sys-
tem.
7-2
Calibrations
Carry out the following calibrations:
Calibrations - Checks
Acquisition Image Chain
Tube Tilt Calibration
Collimator Calibration
Collimation Check
Bad Pixel Calibration
Gain calibration
flat field tests
Half Value Layer
Conversion Factor
AOP: Automatic Optimization of Parameters
AOP Calibration
mAs non linearity
AOP/SNR Check
If AOP/SNR Test not OK, repeat all calibrations from AOP Calibration.
JC-DR-A-266.fm
Service Information and Procedures Class A 2385072-16-8EN
Job Card CAL A027 - Tube Tilt Calibration on page 1681
Job Card CAL A029 - Collimator Format Calibration on page 1683
Job Card ELE A028 - Collimation Checks on page 679
If Collimation Check not OK, repeat Collimator Calibration.
Job Card CAL A041 - Bad Pixel Calibration on page 1693
Job Card CAL A043 - Detector Gain Calibration on page 1697
Job Card ELE A032 – IQ Tools Flat Field Test on page 699
If IQ Tests not OK, repeat all calibrations from Bad Pixel.
Job Card CAL A046 - Half-Value Layer Measurement on page 1703
Job Card CAL A045 - Conversion Factor Measurement on page 1701
Use the same dosimeter as that used for CAL A046
Job Card CAL A051 - AOP Calibration on page 1723
Job Card CAL A053 - mAs Non-linearity Calibration on page 1725
Job Card ELE A038 - AOP Mode and SNR Check on page 707
Launching the Service Desktop on page
Component FW/SW to ADS Version compatibility Matrix
Job Card
Page no. 1367
Senographe DS
Job Card D/R A266 - Tube Tilt Board
120).