GE Healthcare
Revision 1
3. Examine the positioning of the new top cover as follows:
a. Ensure each of the new top cover's sides are firmly in place.
b. Ensure that the back of the new top cover fits firmly in the Bucky framework and that there is equi-
distant space between the top cover and the Bucky.
4. Attach the three screws and brass blocks (4) (cross-head screwdriver) to secure the top cover to the
Bucky. Ensure that you access the two side screws by the hole (5) in the Bucky framework.
! Notice:
Be careful not to overtighten the screws on the three plastic threaded studes of the Bucky. If you
overtighten the screws you could break the new Bucky top cover.
JC-DR-A-258.fm
Service Information and Procedures Class A 2385072-16-8EN
4
Page no. 1351
Job card D/R A258 - Bucky Board
5
Senographe DS
Chapter 9