19. Detach the wrist strap.
20. Replace the SB-150 cover, and use the two screws (cross-head screwdriver) to secure the SB-150
cover to the chassis.
21. Put the SB-150 Workstation back into the Control Station (refer to
Card D/R A311 - AWS
Card D/R A311 - AWS
22. Reinstall the Omega Cover (see
page
503).
23. Reinstall the Top Cover (see
page
503).
24. Reinstall the X-ray Console.
8
COMPLETION
Ensure that the newly installed CD-ROM is correctly configured, as follows:
1. Power on the Senographe system, and login as the sdc user.
2. Insert a blank CD-ROM in the new CD-ROM drive.
3. From the Browser, click the CD-R icon and select Query in the drop-down menu.
4. The blank CD-ROM should be recognized by the Query. If the blank CD-ROM is not recognized,
check connections, and restart the AWS.
JC-DR-A-316.fm
Service Information and Procedures Class A 2385072-16-8EN
Unit), and re-connect it to all the cables (refer to
Unit).
Job Card PHY A040 - Remove/Reinstall Control Station Covers on
Job Card PHY A040 - Remove/Reinstall Control Station Covers on
Page no. 1461
Job Card D/R A316 - SB-150 CD-ROM Drive
Section 6-3-2, page 1407
Section 8, page 1414
Senographe DS
in
Job
in
Job
Chapter 9