GE Senographe DS Manual page 1511

Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
7. If it exists (i.e. on pre-Penduick V3 Control Stations), disconnect the allen screw and bolt (4) connect-
ing the UPS base plate to the L-shaped bracket (2.5 mm allen wrench).
5
5
Penduick and above V3 Control Stations and V4 Control Stations
8. Disconnect the four allen screws (5) connecting the UPS base plate to the Control Station framework
(5 mm allen wrench).
9. Remove the UPS base plate from the Control Station frame to expose the IDC Computer.
10. Disconnect all cables from the IDC Computer rear panel.
JC-DR-A-321_DS.fm
Service Information and Procedures Class A 2385072-16-8EN
4
Pre-Penduick V3 Control Stations
Page no. 1511
Senographe DS
Job Card D/R A321 - IDC Computer
Chapter 9