Revision 1
9-2
Connect the new keyboard
1. Unplug and remove the X-ray Console.
2. Remove Control Station Top Cover. See
Covers on page
503.
3. Remove Control Station Omega Cover. See
Covers on page
503.
4. Disconnect the original keyboard from the back of the ADS Computer, and remove it from the Control
Station.
1
2
5
Reference
1
2
5
3, 4, 6
5. Put the new keyboard on the Control Station and connect it to the USB connector at back of the ADS
Computer (as illustrated above).
6. Reinstall the Omega Cover (see
page
503).
7. Reinstall the Top Cover (see
page
503).
8. Reinstall the X-ray Console.