Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A317 - Radiation Screen
4. Reinstall the edge pinch that you previously removed and secure it with two screws (5).
5
5. Carefully introduce one pinch plate (6) between the edge pinch and the radiation screen.
6
7
6. Tighten with moderation the two screws (7).
7. Install the main frame-radshield joint (8) between the radiation screen and the frame.
8
9
8. Reinstall the eight screw covers (9) to the fixing points of the edge pinches.