Job Card D/R A249 - W304 Cable for Paddle Recognition
6-2
Reassembly of the Paddle Recognition Cable
Reverse the disassembly procedure. Take care to correctly route the cables as shown in the illustration
below.
6-3
Check the Compression Movement
Check that both sets of compression/decompression pedals and the manual compression knobs located
on the paddle holder all function correctly. That is, the paddle moves up or down in response to each
control.
7
COMPLETION
None.