8. Connect the power cable and SCSI cable to the back panel of the new CD-RW unit. The SCSI cable
must be connected to the top SCSI connector (3).
! Notice:
Be careful not to pinch or damage cables.
9. Carefully slide the housing unit that accommodates the new CD-RW unit back into the Control Sta-
tion. Fit, but do not tighten, the two screws and washers (4) under the CD-RW unit (3 mm allen
wrench). The screws should be placed in the second from from the front of the Control Station frame.
Position the CD-RW unit centrally. The right edge of the CD-RW unit should be approximately
12.5 cm from the right lateral bar of the Control Station frame.
10. Temporarily mount the top cover, then center the CD-RW unit's tray with respect to the cover aper-
ture. It is possible and accepted that the width of the CD-RW unit exceeds the width of the opening in
the top cover. However, there should be no gap between the CD-RW and the top cover.
11. Once you have aligned the top cover and the CD-RW unit, remove the top cover and tighen the
screws and washers (4) so that housing accommodating the CD-RW unit is secured to the Control
Station frame.
12. Replace the Control Station top cover (see
Covers on page
503).
7-3
Verification of installation
1. Power up the Senographe system and insert a blank CD-ROM in the new CD-RW unit.
2. From the Browser,click the CD-R icon and select Query in the drop-down menu.
3. The blank CD-ROM should be recognised by the Query. If the blank CD-ROM is not recognized,
check connections, and restart the AWS.
JC-DR-A-314.fm
Service Information and Procedures Class A 2385072-16-8EN
3
Job Card D/R A314 - External CD-RW Unit
4
Job Card PHY A040 - Remove/Reinstall Control Station
Page no. 1441
Senographe DS
12.5 cm
Chapter 9
Need help?
Do you have a question about the Senographe DS and is the answer not in the manual?
Questions and answers
G.E.Essential l have recently bought What is the full weight.of the mamo Full weight of the console and other İtem's also measurements of each item with the main machine please All this Shipping Company required
The GE Senographe DS mammography machine is shipped in three crates with the following dimensions and weights:
- Crate 1 (Gantry, Generator Cabinet, and ramp):
- Dimensions: 2066 mm (81.34 in) depth × 848 mm (33.39 in) width × 2292 mm (90.24 in) height
- Mass: 641 kg (1410.2 lbs)
- Crate 2 (Control Station):
- Dimensions: 1110 mm (43.70 in) depth × 770 mm (30.31 in) width × 1850 mm (72.83 in) height
- Mass: 210.5 kg (464.07 lbs) for truck transport (without side covers and lid)
- Mass: 223.5 kg (492.73 lbs) for plane/boat transport (with side covers and lid)
- Crate 3 (Detector):
- Dimensions: 872 mm (34.33 in) depth × 2270 mm (89.37 in) width × 260 mm (10.24 in) height
- Mass: 104.4 kg (229.68 lbs)
### Total Mass:
- For truck transport: 641 + 210.5 + 104.4 = 955.9 kg (2104.95 lbs)
- For plane/boat transport: 641 + 223.5 + 104.4 = 968.9 kg (2133.61 lbs)
These values include the gantry, generator cabinet, control station, detector, and all related components.