Support (cross-head screwdriver).
2
(Step applicable for Type 3 Compression Potentiometers only)
5. Remove the faulty Compression Potentiometer (3) from the Compression Potentiometer Support.
(Step applicable for Type 3 Compression Potentiometers only)
6. Remove the Allen screws (4) securing the Compression Potentiometer Support (3 mm Allen wrench).
Then remove the Compression Potentiometer Support plus Potentiometer (for Type 1/Type 2) or
Compression Potentiometer Support (Type 3) from the Compression Assembly frame.
4
Type 1
JC-DR-A-243.fm
Service Information and Procedures Class A 2385072-16-8EN
3
4
Job Card D/R A243 - Compression Potentiometer and Cable
Type 2
Page no. 1291
Senographe DS
4
Type 3
Chapter 9