GE Healthcare
Revision 1
Job Card D/R A421 - Conditioner Unit
n
5
PREREQUISITES
Remove Generator Cabinet covers as required for access. See
Generator Covers on page
6
PROCEDURE
6-1
Preparation - Conditioner Drain
To drain the Conditioner, proceed as follows:
1. Shut down the entire system by clicking the System Shutdown button
2. Switch OFF the Generator and MDR circuit breakers in the Generator Cabinet.
3. Prepare the Empty container to receive the used coolant. Place the end of the purge tool into the
container.
4. Disconnect the coolant input pipe
5. Connect the disconnected coolant input pipe to the purge tool. The connector is now unsealed, and
coolant can drain into the prepared container.
6. Switch ON the MDR circuit breaker and press the Conditioner ON button in the generator cabinet.
7. Coolant drains into the container, until the "Check Fluid" alarm stops the flow.
8. Switch OFF the MDR circuit breaker.
Note:
The Conditioner is now empty.
9. Label the container containing the drained coolant to indicate the nature of its contents; cross out the
word Empty on the label, and copy the information from one of the coolant containers (Flush or Top-
ping) provided in the kit. See below for disposal.
Chapter 9
Service Information and Procedures Class A 2385072-16-8EN
513.
from the Conditioner. The pipe connector seals automatically.
Page no. 1634
Job Card PHY A042 - Remove/Reinstall
on the Browser window.
Senographe DS
JC-DR-A-421.fm
Need help?
Do you have a question about the Senographe DS and is the answer not in the manual?
Questions and answers
G.E.Essential l have recently bought What is the full weight.of the mamo Full weight of the console and other İtem's also measurements of each item with the main machine please All this Shipping Company required
The GE Senographe DS mammography machine is shipped in three crates with the following dimensions and weights:
- Crate 1 (Gantry, Generator Cabinet, and ramp):
- Dimensions: 2066 mm (81.34 in) depth × 848 mm (33.39 in) width × 2292 mm (90.24 in) height
- Mass: 641 kg (1410.2 lbs)
- Crate 2 (Control Station):
- Dimensions: 1110 mm (43.70 in) depth × 770 mm (30.31 in) width × 1850 mm (72.83 in) height
- Mass: 210.5 kg (464.07 lbs) for truck transport (without side covers and lid)
- Mass: 223.5 kg (492.73 lbs) for plane/boat transport (with side covers and lid)
- Crate 3 (Detector):
- Dimensions: 872 mm (34.33 in) depth × 2270 mm (89.37 in) width × 260 mm (10.24 in) height
- Mass: 104.4 kg (229.68 lbs)
### Total Mass:
- For truck transport: 641 + 210.5 + 104.4 = 955.9 kg (2104.95 lbs)
- For plane/boat transport: 641 + 223.5 + 104.4 = 968.9 kg (2133.61 lbs)
These values include the gantry, generator cabinet, control station, detector, and all related components.
This answer is automatically generated