GE Healthcare
Revision 1
Job Card D/R A404 - HV Unit
work. The three allen screws are located on the three accessible sides of the HV unit.
CAUTION
Pay particular attention to ensure that you do not damage the kV/mA board when tighten-
ing the allen screw which is in close proximity to the kV/mA board.
3. On the kV/mA board located on the side of the HV unit, make the following cable connections:
•
Gray ribbon cable to the XJ1 connector, ensuring that the red part of the ribbon cable is at the bot-
tom. Be careful when inserting the cable into the XJ1 connector and apply equal pressure to
ensure that you do not bend the connector pins
•
Two black cables into the black connector to the XJ2 connector.
•
The black probe cable with the BNC connector to the XJ3 connector.
Chapter 9
Service Information and Procedures Class A 2385072-16-8EN
1
XJ2
XJ3
XJ1
Page no. 1562
Senographe DS
Note : XJ4 connector is empty
JC-DR-A-404.fm