GE Healthcare
Revision 1
Job Card D/R A404 - HV Unit
work. The three allen screws are located on the three accessible sides of the HV unit.
CAUTION
Pay particular attention to ensure that you do not damage the kV/mA board when tighten-
ing the allen screw which is in close proximity to the kV/mA board.
3. On the kV/mA board located on the side of the HV unit, make the following cable connections:
Gray ribbon cable to the XJ1 connector, ensuring that the red part of the ribbon cable is at the bot-
tom. Be careful when inserting the cable into the XJ1 connector and apply equal pressure to
ensure that you do not bend the connector pins
Two black cables into the black connector to the XJ2 connector.
The black probe cable with the BNC connector to the XJ3 connector.
Chapter 9
Service Information and Procedures Class A 2385072-16-8EN
1
XJ2
XJ3
XJ1
Page no. 1562
Senographe DS
Note : XJ4 connector is empty
JC-DR-A-404.fm