7
PROCEDURE FOR V1/V2 CONTROL STATIONS
7-1
Determine the OpenBoot Version of the SB150 Workstation
1. From behind the Browser, launch a command window (see
page
120).
A command window appears with an ADS prompt.
2. Query the OpenBoot version of the SB150 Workstation, by entering:
prtconf -V
The OpenBoot version of the SB150 Workstation appears.
If the OpenBoot version is lower than 4.10.6, then you cannot change the keyboard with an Inter-
national Keyboard.
If the OpenBoot version is equal to or greater than 4.10.6, then shutdown the Control Station
using the shutdown button the Browser, and proceed to
an International Keyboard.
WARNING
Before proceeding to
Senographe system hospital supply circuit breaker. Lock it in the off position or mark it
clearly with a warning sign to prevent others from applying power.
7-2
Connect the new keyboard
1. Unplug and remove the X-ray Console.
2. Remove Control Station Top Cover. See
Covers on page
503.
3. Remove Control Station Omega Cover. See
Covers on page
503.
4. Disconnect the original keyboard from the back of the ADS Computer, and remove it from the Control
Station.
1
2
3
4
Reference
1
Mouse or Trackball USB Link
2
3
4
Bar Code Reader USB Link
JC-DR-A-315.fm
section
7-2, shut down correctly the equipment and switch off the
Function
Keyboard USB Link
Not used
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A315 - International Keyboard
Launching a Command Line Window on
section 7-2
Job Card PHY A040 - Remove/Reinstall Control Station
Job Card PHY A040 - Remove/Reinstall Control Station
Page no. 1445
Senographe DS
to change the keyboard with