243 (or equivalent medium strength locking compound) to the screws. See picture below.
4. Insert and tighten the two removed lower M4 screws (4) to the corners of the Compression Potenti-
ometer Support to attach the Compression Potentiometer firmly to the Support. Tighten the third
screw also on the upper left corner.
5. Route the Compression Trolley cable W304b (P/N 2379180) and the Potentiometer cable (W303),
JC-DR-A-243.fm
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A243 - Compression Potentiometer and Cable
Page no. 1293
Senographe DS
3
4
Chapter 9