SETUP/GENE/INSTAL/GENE/GENE/BIAS/VOLTAGE/CHANGE
a. Select
sole.
b. Enter 1.300E+2V and valid.
c. Press SETUP/SETUP/SETUP/SETUP/CKSUM/CKSUM on the X-ray Console to perform a
checksum.
d. Press SETUP/SETUP/SETUP/SETUP (i.e. 4 times) on the X-ray Console.
e. Select 30kV/LARGE/MO/MO on the X-ray Console.
f. Do a preparation with the Prep button on the X-ray Console.
g. Check that the result is between 123.5 and 136.5Vdc.
SETUP/GENE/INSTAL/GENE/GENE/BIAS/VOLTAGE/CHANGE
h. Select
sole and enter the value corresponding to the MO/LF noted in the first check point.
7. Fourth check point:
a. Select MO/SF voltage parameter and note the actual value associated to the tube (see step
section 6-2).
b. Enter 2.500E+2V and valid.
c. Press SETUP/SETUP/SETUP/SETUP/CKSUM/CKSUM on the X-ray Console to perform a
checksum.
d. Press SETUP/SETUP/SETUP/SETUP (i.e. 4 times) on the X-ray Console.
e. Select 30kV/SF/MO/MO on the X-ray Console.
f. Do a preparation with the Prep button on the X-ray Console.
g. Check that the result is between 247 and 256Vdc.
8. Fifth check point:
SETUP/GENE/INSTAL/GENE/GENE/BIAS/VOLTAGE/CHANGE
a. Select
sole and enter the value corresponding to the MO/SF noted down on step
b. Select MO/SF voltage parameter and note the default value.
c. Enter 3.000E+2V and valid.
d. Press SETUP/SETUP/SETUP/SETUP/CKSUM/CKSUM on the X-ray Console to perform a
checksum.
e. Press SETUP/SETUP/SETUP/SETUP (i.e. 4 times) on the X-ray Console.
f. Select 30kV/SF/MO/MO on the X-ray Console.
g. Do a preparation with the Prep button on the X-ray Console.
h. Check that the result is between 297 and 305Vdc.
i.
Select SETUP/GENE/INSTAL/GENE/GENE/BIAS/VOLTAGE on the X-ray Console and select
MO/SF.
j.
Enter the value corresponding to the MO/SF value noted down on step
9. If the results are within the tolerances given, the focal bias voltage calibration is finished. If the results
are out of tolerance, repeat the calibration and check procedures.
10. Select SETUP/SETUP/SETUP
values of parameters I, J, K, L, M and N. Each of these six values can be displayed alternately by
rotating the kV dial on the X-ray Console,
11. Delete the previously created patient called "X-ray Tube BIAS cal".
12. Perform a checksum (SETUP/GENE/INSTAL/CKSUM/CKSUM), then turn the system OFF.
13. Remove the high voltage cable adapter. Remember to secure the HV ring correctly using the allen
JC-CAL-A-003.fm
Service Information and Procedures Class A 2385072-16-8EN
Job Card CAL A003 - X-Ray Tube Focal Bias Voltage Scale Factor
/CALIB/PARAM
on the X-ray Console. Note down the calculated
Page no. 1659
Senographe DS
on the X-ray Con-
on the X-ray Con-
on the X-ray Con-
7
(fourth check point).
7
(fourth check point).
Chapter 9
3

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Senographe DS and is the answer not in the manual?

Questions and answers

Dr.Robert Mahmut
March 26, 2025

G.E.Essential l have recently bought What is the full weight.of the mamo Full weight of the console and other İtem's also measurements of each item with the main machine please All this Shipping Company required

1 comments:
Mr. Anderson
March 26, 2025

The GE Senographe DS mammography machine is shipped in three crates with the following dimensions and weights:

- Crate 1 (Gantry, Generator Cabinet, and ramp):
- Dimensions: 2066 mm (81.34 in) depth × 848 mm (33.39 in) width × 2292 mm (90.24 in) height
- Mass: 641 kg (1410.2 lbs)

- Crate 2 (Control Station):
- Dimensions: 1110 mm (43.70 in) depth × 770 mm (30.31 in) width × 1850 mm (72.83 in) height
- Mass: 210.5 kg (464.07 lbs) for truck transport (without side covers and lid)
- Mass: 223.5 kg (492.73 lbs) for plane/boat transport (with side covers and lid)

- Crate 3 (Detector):
- Dimensions: 872 mm (34.33 in) depth × 2270 mm (89.37 in) width × 260 mm (10.24 in) height
- Mass: 104.4 kg (229.68 lbs)

### Total Mass:
- For truck transport: 641 + 210.5 + 104.4 = 955.9 kg (2104.95 lbs)
- For plane/boat transport: 641 + 223.5 + 104.4 = 968.9 kg (2133.61 lbs)

These values include the gantry, generator cabinet, control station, detector, and all related components.

This answer is automatically generated

Table of Contents

Save PDF