Job Card D/R A421 - Conditioner Unit

1
SUPPLIES
Conditioner unit
Coolant kit
2
TOOLS
Standard Tool Box
Purge Tool
3
REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 30 minutes
4
SAFETY PRECAUTIONS
When filling the Conditioner, take great care to keep coolant away from the vents, the ON/OFF switch
and the power cord connection. Do not immerse the Conditioner in any liquid.
Dispose of surplus coolant with care as described in the section Surplus Coolant Disposal; refer to
the Caution below.
CAUTION
• Before adding coolant to the Conditioner, always switch OFF the Generator and MDR cir-
cuit breakers in the Generator Cabinet. Switch on the MDR breaker only as instructed.
• Always use a funnel when adding coolant to the Conditioner.
• Always wear protective gloves and goggles when working with coolant.
• When working with coolant you should treat it with care; avoid unnecessary contact .
See below for First Aid advice:
First Aid in case of contact with Conditioner coolant:
Immediately flush eyes or skin with plenty of water for at least 15 minutes.
If coolant contacts the eye, call a physician immediately.
If coolant is swallowed, drink one to two glasses of water or milk. Avoid alcohol. Call a physician
immediately.
Remove and wash contaminated clothing before reuse.
JC-DR-A-421.fm
Service Information and Procedures Class A 2385072-16-8EN
Page no. 1633
Senographe DS
Job Card D/R A421 - Conditioner Unit