Job Card D/R A331 - UPS
3. Connect the ground cable (3) (small screwdriver) to the TVSS GND connector of the UPS rear panel.
4. Connect all the power cables (4) from the Control Station Components to the power output sockets
on the UPS rear panel, as indicated below:
IDC
AC Multi-
Socket
ADS
Pre-Penduick V3 Control Stations
5. Connect the AC mains supply power cable (5) from to Generator to the power input socket on the
UPS rear panel.
6. Connect the communications cable (6) to the UPS Monitor Port connector of the UPS rear panel.