GE Senographe DS Manual page 1524

Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Job Card D/R A331 - UPS
7
PROCEDURE ON V3/V4 CONTROL STATIONS
7-1
Disassembly of the UPS on V3/V4 Control Stations
1. Locate the UPS unit (1) inside the Control Station.
1
2. Disconnect all cables from the UPS unit rear panel.
Note:
Note locations of disconnected cables.
3. Disconnect the Ground cable (cross-head screwdriver).
4. Open the velcro straps (2) securing the UPS to the Control Station.
Chapter 9
Service Information and Procedures Class A 2385072-16-8EN
2
Page no. 1524
Senographe DS
JC-DR-A-331.fm