GE Healthcare
Revision 1
Job Card D/R A287 - Collimator DC-DC Board
1.
SUPPLIES
Collimator DC-DC Board
2
TOOLS
Standard Tool Box
3
REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 2 hours
4.
SAFETY PRECAUTIONS
WARNING
The Gantry must be powered OFF.
5.
PREREQUISITES
•
Rotate the arm to 180° for access to the collimator.
•
Switch OFF the Gantry power.
•
Remove the tube cover (Refer to
Remove/Reinstall Gantry Covers on page
the Collimator DC-DC Board (1), which is below the Col-
limator Control Board.
JC-DR-A-287.fm
Service Information and Procedures Class A 2385072-16-8EN
Job Card PHY A044 -
523) to access
Page no. 1379
Job Card D/R A287 - Collimator DC-DC Board
Senographe DS
Chapter 9
2
1
3
Chapter 9