e. On V2 Control Stations, thread each of the BNC video cables through the rotative arm support as
shown below.
10. Remove the four screws (11) securing the rotative arm support (5 mm allen wrench) and remove it.
7-2
Reassembly of Sisfle Arm
1. Install the new rotative arm support on the opposite side of the Control Station and secure it with the
same four screws (5 mm allen wrench). To do this easily, use the Sisfle Arm Installation Tool with the
following sub-steps:
a. Put the two threaded rods the Sisfle Arm Installation Tool into the lower two threads (2). Screw
the two threaded rods (1) in a clockwise direction until they no longer move. The right-hand side
JC-DR-A-319.fm
Service Information and Procedures Class A 2385072-16-8EN
11
11