GE Senographe DS Manual page 1677

Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
kV
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
JC-CAL-A-012.fm
Service Information and Procedures Class A 2385072-16-8EN
TABLE 1 - DEFAULT HEATING CURRENT PARAMETERS
LF/Mo
5.19
5.16
5.14
5.13
5.13
5.12
5.12
5.12
5.11
5.08
5.06
5.01
4.99
4.96
4.94
4.92
4.90
4.88
4.86
4.84
4.83
4.82
4.80
4.79
4.77
4.76
4.74
4.73
Job Card CAL A012 - X-Ray Tube Heater Current
SF/Mo
Amperes
4.71
4.73
4.75
4.77
4.79
4.81
4.82
4.83
4.84
4.81
4.78
4.75
4.73
4.71
4.68
4.66
4.64
4.62
4.61
4.59
4.58
4.56
4.55
4.53
4.52
4.51
4.50
4.49
Page no. 1677
Senographe DS
LF/RH
SF/RH
4.87
4.67
4.87
4.68
4.88
4.69
4.88
4.70
4.88
4.72
4.88
4.74
4.88
4.76
4.89
4.78
4.89
4.79
4.87
4.76
4.84
4.74
4.82
4.71
4.80
4.68
4.78
4.66
4.76
4.64
4.74
4.62
4.73
4.60
4.72
4.59
4.70
4.57
4.69
4.56
4.67
4.55
4.66
4.53
4.65
4.52
4.63
4.50
4.62
4.49
4.61
4.48
4.60
4.47
4.59
4.45