Job Card D/R A288 - Collimator Lamp
6
PROCEDURE
Note:
It is not necessary to remove the Collimator when changing the Collimator Lamp.
6-1
Remove the Old Lamp
CAUTION
Wait for the lamp to cool before removing it.
1. Remove the two screws (5, one each side) securing the rear plate (6) which supports the FOV button
(4 mm allen wrench).
6
2. Using an antistatic bag as protection, place the plate and FOV button on the X-ray head support arm.
! Notice:
Do not attempt to disconnect the FOV button.
Do not allow the plate to hang down; the FOV connection is fragile.
3. Remove the four screws (3, two each side) which secure the X-ray containment box (2).
4. Remove the X-ray containment box. If necessary, first remove the two screws (4) which secure the
fan protection box.
5. Remove the four screws (7) securing the right collimator cover (small Pozi screwdriver).
Chapter 9
4
3
5
7
Service Information and Procedures Class A 2385072-16-8EN
2
Page no. 1384
Senographe DS
JC-DR-A-288.fm