GE Senographe DS Manual page 1556

Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

GE Healthcare
Revision 1
Job Card D/R A403 - Capacitor Box
4. Connect the C8+/C7– cable to the new capacitor box with screw S3 (8 mm open-ended wrench).
RC1
S1
S3 (C8+/C7–)
5. Attach the two resonant capacitors, RC1 and RC2, together with their upper plate to the new capaci-
tor box using screws S1 and S2 (10 mm open-ended wrench).
6. Switch on the electrical power supply, both from the Mains Distribution Panel in the room and on the
rotary switch in the Generator cabinet (position I).
Chapter 9
Service Information and Procedures Class A 2385072-16-8EN
Page no. 1556
Senographe DS
RC2
S10
S2
S8
JC-DR-A-403.fm